HISTORICAL VERSIONS

Below is a partial list of standards that have been superseded by newer versions. All NOCSAE standards are continually updated as new data is received.

Historical Versions of ND200

ND200-18

Effective February 2019

Revision

  • Removed COR requirement for baseball projectile Section 13.1.1.

ND200-17am18

Effective June 2018

Revisions

  • Reduced equipment level from 3 to 2 in Section 1.1.
  • Increased pass/fail criteria from 90 lb-f to 112 lb-f in Section 14.1.2.
  • Included a system check for the upper and lower load cells and added an upper range to the allowable load cell response (150 lb.-f) in Section 11.2.

ND0200-21

Effective February 2023

Revisions

  • Eliminated use of the upper and lower load cells and their associated pass/fail criteria
  • Random impact locations updated to impact the cardiac silhouette while shifting the protector instead of moving the NTS.
  • Added definitions for Torso Length and Extent of Protective Coverage
  • Added requirement that manufacturers shall provide the Torso Length of the person the submitted sample is intended to fit.

ND200-19

Effective July 2019

Revision

  • Added requirement to place logo on the primary protective component.

Above is a partial list of standards that have been superseded by newer versions. All NOCSAE standards are continually updated as new data is received.